Operating Basics 2
Operation Manual 21
Correcting Problems
If the GC stops operation because of a fault, check the
display for any messages. Press [Status] and scroll to view
any additional messages.
1 Use the keyboard or data system to stop the alert tone.
Press [Off/No] on the keyboard or turn off the offending
component in the data system.
2 Resolve the problem, for example, by changing gas
cylinders or fixing the leak. See the Troubleshooting
Guide for details.
3 Once the problem is fixed, you may need to either power
cycle the instrument, or use the software keyboard or
data system to turn the problem component off, then on
again. For shutdown errors, you will need to do both.