186 Operation Manual
11 Options
About Options
The [Options] key is used for a group of functions that are
usually set on installation and seldom changed afterward. It
accesses this menu:
Calibration
Press [Calibration] to list the parameters that can be
calibrated. These include:
Inlets
Detectors
ALS
Columns
Oven
Atmospheric pressure
In general, you will only need to calibrate the EPC modules
and capillary columns. ALS, oven, and atmospheric pressure
calibration should only be performed be trained service
personnel.
Maintaining EPC calibration—inlets, detectors, PCM, and AUX
The EPC gas control modules contain flow and/or pressure
sensors that are calibrated at the factory. Sensitivity (slope
of the curve) is quite stable, but zero offset requires periodic
updating.
Flow sensors
The split/splitless and purged packed inlet modules use flow
sensors. If the Auto flow zero feature (see page 186) is on,
they are zeroed automatically after each run. This is the
recommended way. They can also be zeroed manually—see
“To zero a specific flow or pressure sensor.”
Calibration
Communication
Keyboard and Display

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals