Methods and Sequences 4
Operation Manual 39
What Happens When You Load a Method?
There are two kinds of methods:
The active method—This is sometimes referred to as the
current method. The settings defined in this method are
the settings the GC is currently maintaining.
Stored methods—Up to 9 user- created methods can be
stored in the GC, along with one SLEEP method, one
WAKE method, one CONDITION method, an MS VENT
method, and a default method.
When a method is loaded from the GC or Agilent data
system, the setpoints of the active method are immediately
replaced with the setpoints of the method loaded.
The method loaded becomes the active (current) method.
The Not Ready light will stay lit until the GC reaches all
of the settings specified by the method that was just
loaded.
Refer to “Running a Method or a Sequence from the Keypad
for details on using the keypad to load, modify, and save
methods.