EasyManua.ls Logo

Agilent Technologies 7890B - Sulfur Performance

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
82 Operation Manual
6 Chromatographic Checkout
If performing a manual injection (with or without a data
system):
a Press [Prep Run] to prepare the inlet for splitless
injection.
b When the GC becomes ready, inject 1 µL of the
checkout sample and press [Start] on the GC.
c The following chromatogram shows typical results for a
new detector with new consumable parts installed.
Sulfur performance
1 Install the sulfur filter and filter spacer.
2 Ignite the FPD flame if not lit.
3 Display the signal output and monitor. This output
typically runs between 50 and 60 but can be as high as
70. Wait for the output to stabilize. This takes
approximately 1 hour.
If the baseline output is too high:
Check column installation. If installed too high, the
stationery phase burns in the flame and increases
measured output.
Check for leaks.
Bake out the detector and column at 250 °C.
Wrong flows set for installed filter.
Methylparathion
Isooctane

Table of Contents

Other manuals for Agilent Technologies 7890B

Related product manuals