Chromatographic Checkout 6
Operation Manual 55
To Prepare for Chromatographic Checkout
Because of the differences in chromatographic performance
associated with different consumables, Agilent strongly
recommends using the parts listed here for all checkout
tests. Agilent also recommends installing new consumable
parts whenever the quality of the installed ones is not
known. For example, installing a new liner and septum
ensures that they will not contribute any contamination to
the results.
When the GC is delivered from the factory, these consumable
parts are new and do not need replacement.
1 Check the indicators/dates on any gas supply traps.
Replace/recondition expended traps.
2 Install new consumable parts for the inlet and prepare
the correct injector syringe (and needle, as needed).
NOTE
For a new GC, check the installed inlet liner. The liner shipped in the inlet
may not be the liner recommended for checkout.
Table 3 Recommended parts for checkout by inlet type
Recommended part for checkout Part number
Split splitless inlet
Syringe, 10-µL 5181-1267
O-ring 5188-5365
Septum 5183-4757
Liner 5062-3587 or
5181-3316
Multimode inlet
Syringe, 10-µL 5181-1267
O-ring 5188-6405
Septum 5183-4757
Liner 5188-6568
Packed column inlet
Syringe, 10-µL 5181-1267
O-ring 5080-8898

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals