Configuration 10
Operation Manual 167
Analog out 1/Analog out 2
Fast peaks
The GC allows you to output analog data at two speeds. The
faster speed—to be used only with the FID, FPD, and
NPD—allows minimum peak widths of 0.004 minutes (8 Hz
bandwidth), while the standard speed—which can be used
with all detectors— allows minimum peak widths of 0.01
minutes (3.0 Hz bandwidth).
To use fast peaks:
1 Press [Config][Analog out 1] or [Config][Analog out 2].
2 Scroll to Fast peaks and press [On/Yes].
The fast peaks feature does not apply to digital output.
If you are using the fast peaks feature, your integrator must be
fast enough to process data coming from the GC. Integrator
bandwidth should be at least 15 Hz.