EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
Chromatographic Checkout 6
Operation Manual 89
7
The following chromatogram shows typical results for a
new detector with new consumable parts installed.
Sulfur performance
1 Install the sulfur filter.
2 Ignite the FPD flame, if not lit.
3 Display the signal output and monitor. This output
typically runs between 50 and 60 but can be as high as
70. Wait for the output to stabilize. This takes
approximately 2 hours.
If the baseline output is too high:
Check column installation. If installed too high, the
stationery phase burns in the flame and increases
measured output.
Check for leaks.
Bake out the detector and column at 250 °C.
Wrong flows set for installed filter
If the baseline output is zero, verify the electrometer is on
and the flame is lit.
4 If using a data system, prepare the data system to
perform one run using the loaded checkout method. Make
sure the data system will output a chromatogram.
Tributylphosphate
Isooctane t-Butyldisulfide

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals