Methods and Sequences 4
Operation Manual 47
To load a stored sequence
1 Press [Load][Seq].
2 Enter the number of the sequence to be loaded (1–9).
3 Press [On/Yes] to load the sequence or [Off/No] to cancel
the load.
An error message is displayed if the specified sequence
number has not been stored.
To determine sequence status
Press [Seq Control] to display the current status of the active
sequence. There are six possible sequence status modes:
• Start/running
• Ready wait
• Paused/resume
• Stopped
• Aborted
• No sequence
Automating Data Analysis, Method Development, and Sequence
Development
The output of the detectors is digitized and can be sent to
an automated data analysis system (such as Agilent
OpenLAB CDS), where it is analyzed and the results
summarized in reports.
The Agilent data system also can be used to create and store
methods and sequences that are sent to the GC through a
network.