128 Operation Manual
8 Early Maintenance Feedback
To Reset an EMF Counter
When a Service Due counter passes its threshold, the GC
Service Due indicator lights.
1 Press [Service Mode].
2 Scroll to Maintenance and press [Enter].
3 Each EMF component with a counter that passed its
threshold will be marked with an asterisk. Scroll to the
desired GC component (front or back inlet, front or back
detector, valve, instrument, and so forth) and press [Enter]
to select it. The list of counters for this component
display. Each component that passed its threshold will be
marked with an asterisk.
4 Scroll to the desired counter.
5 Press [Off/No] to reset the counter to 0.