Configuration 10
Operation Manual 141
Aux EPC 7,8,9
Status
Time
Valve #
2 Dimensional GC Valve
Front injector
Back injector
Sample tray
Instrument
In many cases you can move directly to the item of
interest by pressing [Config][device].
3 When the Configure Device Display opens, the cursor
should be on the Unconfigured field. Press [Mode/Type] and
follow the GC prompts to assign resources to the device.
4 After assigning resources, the GC prompts for you to
power cycle the GC. Turn the GC power switch off and
then on.
When the GC starts, select the device just assigned the GC
resources for further configuration if needed. When accessed,
its mode should indicate Configured and the other
configuration properties are displayed.
Setting configuration properties
A device’s configuration properties are constant for an
instrument hardware setup unlike method settings which can
change from sample run to sample run. Two example
configuration settings are the gas type flowing through a
pneumatic device and the operation temperate limit of a
device.
To change the setting configuration properties for a
Configured device:
1 Press [Config] on the GC keypad and select a device from
the list, then press [Enter].
In many cases you can move directly to the item of
interest by pressing [Config][device].
2 Scroll to the device setting and change the property. This
can involve making a selection from a list using
[Mode/Type], using [On/Yes] or [Off/No], or entering a
numeric value. Press [Info] for help on changing numeric
settings, or see the section of this document describing
the specific configuration of the device.