EasyManuals Logo

Agilent Technologies 7890B User Manual

Agilent Technologies 7890B
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
164 Operation Manual
10 Configuration
3
Scroll to Cryo Type (Valve BV).
If the value is not N2, press [Mode/Type], select N2 Cryo,
press [Enter] and then [Clear].
This tells the GC what coolant will be used.
Configure the user-configurable heater
Many of the following steps tell you to reboot the GC. Ignore
these requests by pressing [Clear]. Do not reboot until
specifically told to do so in these instructions.
1 Press [Config] and select Aux 1. Press [Enter].
2 Enter the following control values. Press [Enter], then
[Clear] after each one.
a Proportional Gain—5.30
b Integral Time—10
c Derivative Time—1.00
d Mass (Watt- sec/deg)—18
e Power (Watts)—To find the watts to set here, scroll to
Back Inlet Status (BINLET). Note the watts value and enter
it for this parameter.
f Cryo Control Mode—Press [Mode/Type]. The first line
should already be PTV. Select Cryo Trap.
g Zone Control mode—Press [Mode/Type] and select PTV.
h Sensor—Press [Mode/Type] and select Thermocouple.
i Maximum Setpoint—400
j Maximum Programming Rate—720
Reboot the GC
Press [Options], select Communications, and press [Enter]. Select
Reboot GC and press [On/Yes] twice.

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals