192 Operation Manual
11 Options
10
Scroll to Measured flow and enter the corrected column
flow rate (in mL/min) from the run performed above.
11 Scroll to Unretained peak and enter the actual elution time
from the run performed above.
12 When you press [Enter], the GC will estimate the column
length or diameter based on the elution time input and
will now use that data for all calculations.