Chromatographic Checkout 6
Operation Manual 61
If performing a manual injection (with or without a data
system):
a Press [Prep Run] to prepare the inlet for splitless
injection.
b When the GC becomes ready, inject 1 µL of the
checkout sample and press [Start] on the GC.
c The following chromatogram shows typical results for a
new detector with new consumable parts installed and
nitrogen makeup gas.
mi
012345
pA
0
50
100
150
200
250
300
350
400
FID1 A, (C:\FID.D)
C13 C14
C15 C16