Chromatographic Checkout 6
Operation Manual 79
Septum purge 15 mL/min
PTV inlet
Mode Splitless
Inlet temperature 75 °C
Initial time 0.1 min
Rate 1 720 °C/min
Final temp 1 350 °C
Final time 1 2 min
Rate 2 100 °C/min
Final temp 2 250 °C
Final time 2 0 min
Purge time 0.75 min
Purge flow 60 mL/min
Septum purge 3 mL/min
Detector
Temperature 200 °C (On)
Hydrogen flow 60 mL/min (On)
Air (Oxidizer) flow 60 mL/min (On)
Mode Constant makeup flow OFF
Makeup flow 60 mL/min (On)
Makeup gas type Nitrogen
Flame On
Lit offset Typically 2 pA
PMT voltage On
Emission Block 125 °C
Oven
Initial temp 70 °C
Initial time 0 min
Rate 1 25 °C/min
Final temp 1 150 °C
Final time 1 0 min
Rate 2 5 °C/min
Table 8 FPD
+
Checkout Conditions (continued)(P)