6-8 Service Guide
Digital Control Troubleshooting 8719ET/20ET/22ET
Display Troubleshooting (A18,A19) 8719ES/20ES/22ES
RF Network Analyzers
Dark Pixels Specifications
Dark “stuck on” pixels may appear against a white background. To test for these dots,
press the following:
In a properly working display, the following will not occur:
more than 12 stuck pixels (not to exceed a maximum of 7 red, green, or blue)
more than one occurrence of 2 consecutive stuck pixels
stuck pixels less than 6.5 mm apart
Newtons Rings
To check for the patterns known as Newtons Rings, change the display to white by pressing
the following keys:
Figure 6-3 illustrates acceptable and non-acceptable examples of Newtons Rings.
Figure 6-3 Newtons Rings
System
SERVICE MENU
TESTS
62 x1
EXECUTE TEST
CONTINUE
System
SERVICE MENU
TESTS
62
x1
EXECUTE TEST
CONTINUE

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals