11-16 Service Guide
Error Terms 8719ET/20ET/22ET
Error Term Descriptions 8719ES/20ES/22ES
RF Network Analyzers
Transmission Tracking (ETF and ETR)
Transmission tracking is the difference between the frequency response of the reference
path (including R input) and the frequency response of the transmission test path
(including A or B input) while measuring transmission. The response of the test port cables
is included. These terms are characterized by measuring the transmission (S21, S12) of the
“thru” configuration during the measurement calibration. Note that coupler response is
included in this error term. Typically, this appears as a roll-off below 500 MHz, which is
typically 18 dB at 100 MHz, and 25 dB at 50 MHz
Significant System Components. Large variations in this error term probably indicate
a problem in the reference signal path (if both ETF and ETR are bad) or in the A or B input
path. The “thru” cable also has an effect on transmission tracking.
Affected Measurements. All transmission measurements are affected by transmission
tracking errors.
Figure 11-6 Typical ETF/ETR

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals