2-30 Service Guide
System Verification and Performance Tests 8719ET/20ET/22ET
System Verification 8719ES/20ES/22ES
RF Network Analyzers
If the System Fails the Verification Test
Disconnect and reconnect the device that failed the verification. Then measure the
device again.
If the performance verification still fails:
Continue to measure the rest of the verification devices and printout the results of all
four measurement parameters.
Print the error terms and examine them for anomalies near the failure frequencies.
Refer to Chapter 11 , “Error Terms.”
Make another measurement calibration and follow the flow chart in Figure 2-14.
Figure 2-14 Verification Fails Flowchart