EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8720E Series

Agilent Technologies 8720E Series Service Guide

Agilent Technologies 8720E Series
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #225 background imageLoading...
Page #225 background image
Service Guide 7-17
8719ET/20ET/22ET Source Troubleshooting
8719ES/20ES/22ES Broadband Phase Lock Problems
RF Network Analyzers
4. Turn on the analyzer.
5. To set the internal counter to the frac-N node, press the following:
6. In turn, jumper each of the three supply voltages to A14TP14 and observe the frequency
as shown in Table 7-2. A variation of ±10% is expected. The counter frequency needs to
be extremely out of specification to constitute a failure.
7. If the frequency changes are not correct, replace A14.
8. If the frequency changes are correct, continue with “Divide-by-N (A14) Circuit Check”
next.
Divide-by-N (A14) Circuit Check
NOTE The A13 assembly should still be out of the instrument and the A14 assembly
on an extender board.
1. Jumper A14TP14 to the +5VU supply.
2. Connect an oscilloscope to A14J3 (labeled VCO/N OUT).
3. Press .
4. Vary the fractional-N frequency from 120 MHz to 242 MHz.
If the period of the signal does not vary from 7.5 µsec to 15.5 µsec, replace A14. A
variation of ±10% is expected. The counter frequency needs to be extremely out of
specification to constitute a failure.
If the period does vary as prescribed, remove the jumper and reinsert A14. Always
turn off the analyzer before removing or replacing any assembly.
A14-to-A13 Digital Control Signals Check
1. Place A13 on the extender board and reconnect all of the flexible cables (the one to
A14J1 is optional).
2. The A14 assembly generates a TTL cycle start (CST) signal every 10 microseconds
when the VCO is oscillating.
Table 7-2 VCO Exercise Matrix
Supply Test
Point
Voltage
Mnemonic
A14 Test Point Counter
Frequency
A8TP7 +15V A14TP14 240 MHz
A8TP8 +5VU A14TP14 155 MHz
A8TP2 AGND A14TP14 105 MHz
System
SERVICE MENU
ANALOG BUS ON
Meas
ANALOG IN Aux Input
COUNTER: FRAC-N
System
SERVICE MENU
SERVICE MODES
FRACN TUNE ON

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals