Service Guide 11-5
8719ET/20ET/22ET Error Terms
8719ES/20ES/22ES Error Term Inspection
RF Network Analyzers
Error Term Inspection
Table 11-1 (ES models) and Table 11-2 (ET models) list the calibration coefficients along
with their corresponding test numbers. Refer to the appropriate table when performing
this procedure.
Table 11-1 Calibration Coefficient Terms and Tests for ES Models
Calibration
Coefficient
Calibration Type Test Number
Response Response and
Isolation*
1-port 2-port
1E
R
or E
T
E
X
(E
D
)E
D
E
DF
31
2E
T
(E
R
)E
S
E
SF
32
3E
R
E
RF
33
4E
XF
34
5E
LF
35
6E
TF
36
7E
DR
37
8E
SR
38
9E
RR
39
10 E
XR
40
11 E
LR
41
12 E
TR
42
NOTES:
Meaning of first subscript: D=directivity; S=source match; R=reflection tracking;
X=crosstalk; L=load match; T=transmission tracking.
Meaning of second subscript: F=forward; R=reverse.
* Response and Isolation cal yields: E
X
or E
T
if a transmission parameter (S
21
, S
12
) or E
D
or
E
R
if a reflection parameter (S
11
, S
22
)

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals