14-12 Service Guide
Assembly Replacement and Post-Repair Procedures 8719ET/20ET/22ET
Removing the Display Lamp and Assembly 8719ES/20ES/22ES
RF Network Analyzers
Removing the Display Lamp and Assembly
Tools Required
T-10 TORX screwdriver
T-15 TORX screwdriver
small slot screwdriver
ESD (electrostatic discharge) grounding wrist strap
5/16-inch open-end torque wrench (set to 10 in-lb)
Removal
1. Remove the front panel assembly (A1), refer to “Removing the Front Panel Assembly”
on page 14-8.
2. Remove the 2 ribbon cables between:
(Item 1) the inverter assembly (A20) and the front panel interface board (A2)
(Item 2) the inverter assembly (A20) and the display assembly (A18)
3. Remove the 3 screws (item 3) that attach the display cover to the front panel.
4. Lift the display from the front panel and remove the 2 screws (item 4) from the outside
of the display.
5. Pull the lamp (item 5) out with a curving side motion, as shown.
Replacement
1. Reverse the order of the removal procedure.

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals