3-14 Service Guide
Adjustments and Correction Constants 8719ET/20ET/22ET
Protected Option Numbers Correction Constants (Test 50) 8719ES/20ES/22ES
RF Network Analyzers
6. Press the following:
7. Observe the analyzer for the results of the adjustment routine:
If the analyzer displays Option Cor DONE, you have completed this procedure.
8. If no more correction constant routines are going to be performed, return the A7 Switch
to the NRM position and perform the EEPROM Backup Disk Procedure,” on page 3-33.
In Case of Difficulty
If the analyzer displays Option Cor FAIL, check the keyword used in step 4 and make
sure it is correct. Pay special attention to the letters “I” or “O”, the numbers “1” or “0”.
Repeat this entire adjustment test.
If the analyzer continues to fail the adjustment routine, refer to Chapter 6 , “Digital
Control Troubleshooting,” or contact the nearest Agilent Technologies sales or service
office.
System
SERVICE MENU
TESTS
50 x1
EXECUTE TEST
YES

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals