Service Guide 3-23
8719ET/20ET/22ET Adjustments and Correction Constants
8719ES/20ES/22ES Output Power Adjustments
RF Network Analyzers
In Case of Difficulty
With the power meter connected to PORT 1 (REFLECTION port on the ET models),
check for available power and modulator functionality by adjusting the power DAC
from 0 to 4095.
The power should vary from 20 dB to +4 dB from the maximum specified power in
most cases. (At some of the frequencies you may not be able to tune above maximum
power.)
If your analyzer is not operating correctly, as indicated from the results of the previous
step, refer to Chapter 7 , “Source Troubleshooting,” or contact the nearest Agilent
Technologies sales or service office.