EasyManua.ls Logo

Agilent Technologies Agilent 7890 Series

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the COC Inlet 7
Maintaining Your GC 115
6
Remove the existing insert from the inlet by gently
pushing it out from below with a wire or piece of column.
Store the insert for possible later use.
7 Check that the insert is the correct size for both the
needle and column. (See “To Check the Needle- to- Column
Size on the COC Inlet” on page 110.)
8 Lower the new insert straight into the inlet from the top.
The insert can be installed either end up.
9 Install the spring on top of the insert.
10 Install the septum and septum nut or duckbill septum and
cooling tower assembly and finger- tighten.
11 Install the column. (See “To Install a Capillary Column
with the COC Inleton page 107.)

Table of Contents

Related product manuals