Maintaining the PTV Inlet 9
Maintaining Your GC 155
To Install a Capillary Column with the PTV Inlet
1 Gather the following (see “Consumables and Parts for the
PTV Inlet” on page 152):
Column
Graphpak 2M ferrule
Column nut
Column cutter
Septum
Isopropanol
Lab tissue
Metric ruler
5- mm and 6- mm wrenches
Typewriter correction fluid or a marker
Lint- free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Before installing the column, install the correct column
adapter. (See “To Replace the Inlet Adapter for the PTV
Inlet” on page 170.)
4 Place the column on the hanger with the ends pointing up
and the label to the front.
5 Place a Graphpak 2M ferrule onto the column, with the
graphite facing up towards the inlet.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals