Maintaining the Purged Packed Inlet 6
Maintaining Your GC 81
To Install an Adapter on the Purged Packed Inlet
1 Gather the following:
Brass tubing nut (see “Consumables and Parts for the
Purged Packed Inlet” on page 68.)
Adapter (0.53 mm, 1/8- inch packed, or 1/4- inch
packed)
7/16- inch and 9/16- inch wrench
Vespel/graphite ferrule
Methanol
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Clean the end of the adapter with a lint- free cloth and
methanol to remove contamination such as fingerprints.
4 Place the tubing nut and Vespel/graphite ferrule on the
adapter.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.