84 Maintaining Your GC
6 Maintaining the Purged Packed Inlet
To Change the Glass Liner on the Purged Packed Inlet
1 Gather the following:
Replacement glass liner (See “Consumables and Parts
for the Purged Packed Inlet” on page 68.)
9/16- inch wrench
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Loosen the top insert weldment to remove the top portion
of the inlet.
4 Use a thin wire or wood splint to carefully lift and
remove the old glass liner.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Top insert weldment
Flared end
Glass liner
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.