EasyManuals Logo

Agilent Technologies Agilent 7890 Series User Manual

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
84 Maintaining Your GC
6 Maintaining the Purged Packed Inlet
To Change the Glass Liner on the Purged Packed Inlet
1 Gather the following:
Replacement glass liner (See “Consumables and Parts
for the Purged Packed Inlet” on page 68.)
9/16- inch wrench
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Loosen the top insert weldment to remove the top portion
of the inlet.
4 Use a thin wire or wood splint to carefully lift and
remove the old glass liner.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Top insert weldment
Flared end
Glass liner
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals