64 Maintaining Your GC
5 Maintaining the Split/Splitless Inlet
To Clean the Split/Splitless Inlet
1 Gather the following:
Replacement septum (See “Consumables and Parts for
the Split/Splitless Inlet” on page 44.)
Replacement liner
Replacement O- ring
Replacement gold seal
Replacement washer
Solvent that will clean the type of deposits in your
inlet
Compressed, filtered, dry air or nitrogen
Beaker
Cleaning brushes—The FID cleaning kit (part number
9301- 0985) contains appropriate brushes
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Remove the inlet liner. (See “To Change the Liner and
O- Ring on the Split/Splitless Inlet” on page 56.)
4 Disconnect the column from the inlet.
5 Remove the reducing nut and gold seal. (See “To Replace
the Gold Seal on the Split/Splitless Inlet” on page 59.)
6 Place a beaker in the oven under the inlet to catch the
solvent.
7 Soak a cleaning brush in the solvent and scrub the inside
of the inlet weldment. Repeat 10 times.
8 Rinse the inlet with the solvent.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals