EasyManua.ls Logo

Agilent Technologies Agilent 7890 Series

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
302 Maintaining Your GC
15 Maintaining the FPD
+
To Replace the FPD
+
Ignitor
1 Gather the following:
Ignitor replacement kit. (See “Consumables and Parts
for the FPD
+
on page 290.)
Torx screwdrivers, T- 20 and T- 10
5/16- inch nut driver (or wrench)
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Remove the FPD cover. (See “To Remove the FPD
+
Cover”
on page 300.)
4 Loosen the collet screw holding the cable assembly to the
ignitor. Remove the collet and cable assembly.
5 Use a nut driver to loosen and remove the glow plug.
If using a 5/16- inch wrench, you may need to remove
the exit tube assembly using a T- 10 Torx driver
WARNING
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.

Table of Contents

Related product manuals