EasyManua.ls Logo

Agilent Technologies Agilent 7890 Series

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the NPD 14
Maintaining Your GC 285
To Check the NPD Leakage Current
1 Load the analytical method.
2 Set the NPD Adjust Offset to Off and the Bead Voltage to
0.00 V.
Leave the NPD at operating temperature
Leave flows on or off
3 Press [Front Detector] or [Back Detector], then scroll to
Output.
4 Verify that the output (leakage current) is stable and
< 2.0 pA.
The output should slowly drop towards 0.0 pA, and
should stabilize in the tenths of a picoamp. Current
> 2.0pA indicates a problem.

Table of Contents

Related product manuals