Maintaining the FID 11
Maintaining Your GC 203
To Attach a Capillary Column Adapter on an Adaptable FID
1 Gather the following materials:
Adapter (See “Consumables and Parts for the FID” on
page 196.)
1/4- inch brass nut
1/4- inch Vespel/graphite ferrule
Column cutter
1/4- inch wrench
9/16-inch open-end wrench
Lint- free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Assemble the nut and ferrule onto the adapter.
4 Insert the adapter straight into the detector base as far as
possible.
5 Hold the adapter in this position and finger- tighten the
nut.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
1/4-inch ferrule
1/4-inch nut
Adapter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals