Maintaining the FID 11
Maintaining Your GC 205
To Install a Capillary Column in the FID
1 Gather the following materials (see “Consumables and
Parts for the FIDon page 196.):
Column
Ferrule(s)
Column nut
Column cutter
1/4- inch open- end wrench
Septum
Isopropanol
Lab tissue
Lint- free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 If using the adaptable detector, verify that the adapter is
installed. (See “To Attach a Capillary Column Adapter on
an Adaptable FID” on page 203.)
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Capillary optimized fitting
Adaptable fitting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals