Maintaining the MMI 8
Maintaining Your GC 137
To Change the Septum on the MMI
1 Gather the following:
Replacement septum. (See “Consumables and Parts for
the MMI” on page 128.)
Wrench, hex for changing septum
0- or 00- grade steel wool (optional)
Tweezers
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Remove the septum retainer nut or Merlin cap.
4 Use tweezers to remove the septum or Merlin Microseal
from the insert assembly. Do not gouge or scratch the
interior of the insert assembly.
5 Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal
should face down (toward the oven).
6 Install the septum retainer nut or Merlin cap and
finger- tighten. Tighten the septum retainer nut until the
C- ring is about 1 mm above the nut.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Septum retainer nut
Septum
Insert assembly

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals