Maintaining the FPD
+
15
Maintaining Your GC 293
To Install a Packed Column Adapter in the FPD
+
This procedure applies to 1/8- inch packed metal columns.
1 Gather the following:
FPD packed column adapter (See “Consumables and
Parts for the FPD
+
on page 290.)
Polyimide ferrule for 530- µm column
7/16- inch wrench
5/16- inch wrench
1/8- inch nut and ferrule for a 1/8- inch packed column
Adapter nut
Lint- free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Place a 1/8- inch nut and ferrule onto the packed column.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Adapter ferrule
Adapter nut
1/8-inch Packed
column adapter
1/8-inch Packed column
Ferrule
1/8-inch Nut
Capillary tube
'HWHFWRUÀWWLQJ

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals