Maintaining the VI 10
Maintaining Your GC 185
To Remove the VI Interface
1 Gather the following:
• 1/4- inch and 7- mm wrench
• T- 20 Torx screwdriver
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Remove the column.
4 Remove the transfer line by loosening the nut, then lift it
from the interface.
5 Loosen the five inlet cover screws and remove the inlet
cover.
6 Remove the clamping plate from the interface by
loosening the captive screw.
7 Lift the interface out of the heater block.
WARNING
Be careful! The interface may be hot enough to cause burns. If the
interface is hot, wear gloves to protect your hands.