Maintaining the uECD 13
Maintaining Your GC 251
To Install a Capillary Column in the uECD
1 Gather the following:
Ferrule (See “Consumables and Parts for the uECD” on
page 244.)
Column nut
Septum
Column
1/4- inch, 5/16- inch, and 9/16- inch wrenches
Column cutter
Lint- free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Load the inlet maintenance method and wait for the GC
to become ready.
4 Place a septum (if the column id is 0.2 mm), capillary
column nut, and ferrule on the column.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals