116 Maintaining Your GC
7 Maintaining the COC Inlet
To Clean the COC Inlet
1 Gather the following:
• 1/4- inch and 5/16- inch wrenches
• Cleansing bath
• Aqueous detergent
• Distilled water
• Methanol
• Compressed, filtered, dry air or nitrogen
• Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Remove the column.
4 Remove the septum nut or cooling tower and then remove
the septum.
5 Remove the existing insert from the inlet by gently
pushing it out from below with a wire or piece of column.
Store the insert for possible later use.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals