92 Maintaining Your GC
6 Maintaining the Purged Packed Inlet
To Install a Packed Column Adapter on a Detector Fitting
1 Gather the following:
7/16- inch, 9/16- inch, and 1/2- inch wrenches
Vespel/graphite ferrule (See “Consumables and Parts
for the Purged Packed Inlet” on page 68.)
Brass column nut
Lint- free gloves
Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
Consumables and Parts for the FID (Packed columns
can only be installed on an adaptable FID.)
Consumables and Parts for the TCD
Consumables and Parts for the NPD
Consumables and Parts for the FPD
For uECD, 1/4- inch packed columns require no
adapter. For 1/8- inch packed columns, see
Consumables and Parts for the uECD.
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Assemble a nut and a ferrule onto the adapter.
WARNING
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals