Maintaining the PTV Inlet 9
Maintaining Your GC 161
To Replace the Septumless Head PTFE Ferrule on the PTV Inlet
1 Gather the following:
Syringe with 23- gauge needle (See “Consumables and
Parts for the PTV Inleton page 152.)
Replacement PTFE ferrule
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Unscrew the guide cap from the septumless head and
remove the PTFE ferrule.
4 Push the guide cap and replacement ferrule over the
syringe needle so that at least 10 mm of the needle tip is
exposed.
5 Guide the end of the needle into the septumless head
until the ferrule meets the septumless head.
6 Loosely install the guide cap.
7 If not configured, configure the column.
8 Put the inlet into Splitless mode.
9 Set the column flow to 5 mL/min and the purge flow to
60 mL/min.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals