Maintaining the PTV Inlet 9
Maintaining Your GC 167
To Change the Liner on the PTV Inlet
1 Gather the following:
Installation tool for 3D ferrules (See “Consumables and
Parts for the PTV Inleton page 152.)
Assembly tool (part number G2617- 80540)
Replacement liner
Graphpak 3D ferrule
5/16- inch wrench
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Remove the head from the inlet:
For septumless head, disconnect the carrier gas line
and unscrew and remove the septumless head assembly
from the inlet.
For septum head, loosen the septum head assembly
from the inlet. Lift the head to clear the inlet and push
to either side. Do not bend the 1/16- inch lines too
much.
4 Grasp the liner by the Graphpak 3D ferrule. Remove the
liner from the inlet.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals