Maintaining the VI 10
Maintaining Your GC 181
To Install a Capillary Column with the VI
1 Gather the following:
Long (65 mm) column nut
Column
Ferrule
Column cutter
Isopropanol
Septum
1/4- inch wrench
Lab tissue
Metric ruler
Lint- free gloves
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Place the column on the hanger with the ends pointing up
and the label to the front.
4 Place a septum, long capillary column nut, and ferrule on
the column.
Use a long column nut. (See “Consumables and Parts for
the VI” on page 178.)
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals