Maintaining Capillary Columns 4
Maintaining Your GC 37
To Cut a Loop from a Column
1 Gather the following:
New ferrule(s) for the column inlet connection
Column cutter
2 Load the inlet maintenance method and wait for the GC
to become ready.
3 Loosen the inlet column nut and remove the column from
the inlet.
4 Uncoil one loop of column from the column hanger.
5 Cut the unwanted loop from the column.
6 Install the column into the inlet using the new ferrules.
See:
To Install a Capillary Column with the Split/Splitless
Inlet
To Install a Capillary Column with the Multimode Inlet
To Install a Capillary Column with the Purged Packed
Inlet
To Install a Capillary Column with the COC Inlet
To Install a Capillary Column with the PTV Inlet
To Install a Capillary Column with the VI
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.