Maintaining the Purged Packed Inlet 6
Maintaining Your GC 99
To Install Ferrules on a Packed Metal Column
1 Gather the following:
Wrenches
Stainless steel male Swagelok fitting, 1/4- or 1/8- inch
od
Brass Swagelok nut and ferrule set (See “Consumables
and Parts for the Purged Packed Inlet” on page 68.)
Lint- free gloves
2 Verify that the column end is cut square and is free of
burns and deformation.
3 Secure the fitting in a bench vise.
4 Assemble a Swagelok nut and ferrules onto the column.
5 Fully insert the column into the vise- held fitting, then
withdraw 1–2 mm. Finger- tighten the nut.
6 Tighten the nut an additional 3/4 turn with a wrench (for
a 1/8- inch column) or an additional 1- 1/4 turn (for a
1/4- inch column).
7 Unscrew the column nut from the vise- held fitting and
remove the column. Ferrules should now be set in place
on the column with the column end correctly positioned.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Front ferrule
Back ferrule
Column nut

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals