98 Maintaining Your GC
6 Maintaining the Purged Packed Inlet
is never higher than the maximum temperature limit for
the column; 30 °C less than the maximum is usually
sufficient.
7 Continue conditioning overnight at the final temperature.
Cool the oven to room temperature with carrier flow on.
8 Attach the column to the detector and maintain
established flow. (See “To Install a Packed Metal Column”
on page 90.)