162 Maintaining Your GC
9 Maintaining the PTV Inlet
10
After the inlet pressurizes, press [Prep Run] twice.
11 Observe the inlet Tota l Flo w. Tighten the guide cap until
the Tot a l Fl ow stops decreasing (typically near 8 mL/min).
12 Remove the syringe from the inlet and press [Stop].
13 Restore the analytical method.
14 Reset the EMF counter. See To Reset an EMF Counter in
the Operation Manual.