EasyManuals Logo

Agilent Technologies Agilent 7890 Series User Manual

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #284 background imageLoading...
Page #284 background image
284 Maintaining Your GC
14 Maintaining the NPD
26
Install the bead assembly and restore normal operating
conditions. (See “To Replace the NPD Bead Assembly” on
page 272.) (Do not reset the bead counter unless
replaced.)
After installing new collector parts, the NPD leakage
current should be lower. (See “To Check the NPD Leakage
Current” on page 285.) If the leakage current is abnormal,
check for proper reassembly of the detector (especially
where the electrometer interconnect contacts the collector
assembly) and for leaks.
27 Reset the EMF counters. See To Reset an EMF Counter in
the Operation Manual.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals