EasyManua.ls Logo

Agilent Technologies Agilent 7890 Series

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the Split/Splitless Inlet 5
Maintaining Your GC 65
9
Blow the inside of the inlet dry with compressed air or
nitrogen.
10 Install the gold seal and reducing nut.
11 Install the liner and O- ring.
12 Install the column. (See “To Install a Capillary Column
with the Split/Splitless Inlet” on page 48.)
13 Check for leaks.
14 Bakeout contaminants. (See “To Bakeout Contaminants
from the Split/Splitless Inlet” on page 66.)
15 Restore the analytical method.
16 Reset the septum and liner EMF counters. See To Reset
an EMF Counter in the Operation Manual.

Table of Contents

Related product manuals