Maintaining the Purged Packed Inlet 6
Maintaining Your GC 85
5
Wearing gloves, inspect the O- ring and replace, if
necessary. (See “To Change the O- Ring on the Purged
Packed Inleton page 83.)
6 Wearing gloves, grasp the flared end (top) of the
replacement glass liner with tweezers and install it in the
inlet. If the glass liner does not seat properly because a
capillary column is installed, remove the column, install
the glass liner, and replace the column. (See “To Install a
Capillary Column with the Purged Packed Inlet” on
page 72.)
7 Install the top insert weldment and hand- tighten firmly.
8 Restore the analytical method.
9 Configure the new liner. If using the optional barcode
scanner, scan the liner directly into your data system
software configuration, then download the changes to the
GC.
10 Reset the EMF counter. See To Reset an EMF Counter in
the Operation Manual.