124 Maintaining Your GC
7 Maintaining the COC Inlet
6
Loosen the retaining nut just enough so the needle is
again free.
7 Depress the syringe plunger slowly until it pushes the
needle to the end of the barrel, then finger- tighten the
retaining nut.
8 Use a solvent to rinse the syringe and check for leaks or
blocks. Leaks may be fixed by further tightening the
retaining nut. Blocks or serious leaks require repeating
this procedure.