Maintaining the PTV Inlet 9
Maintaining Your GC 171
5
Remove the inlet adapter and discard the old silver seal.
6 Insert a new silver seal into the adapter and
finger- tighten the adapter onto the inlet. Tighten an
additional 1/16 to 1/8 turn with a wrench; overtightening
will damage the inlet.
7 Install the column. (See “To Install a Capillary Column
with the PTV Inlet” on page 155.)
8 Check the adapter for leaks.
9 Restore the analytical method.
10 Reset the silver seal EMF counter. See To Reset an EMF
Counter in the Operation Manual.
Silver seal
Graphpak inlet adapter
Column nut
Ferrule for Graphpak adapter