Page #221 background imageLoading...
Page #221 background image
Maintaining the FID 11
Maintaining Your GC 221
c
Insert the other insulator onto the top of the collector,
with the flat surface facing towards the housing.
15 Install the FID ignitor castle.
a Orient the castle so that the threaded hole for the
ignitor faces toward the electronics.
b Insert the FID castle into the collector housing.
c Install the spring washer over the castle.